Inicio / News / What are the requirements for the installation site of the flowmeter? 

What are the requirements for the installation site of the flowmeter?  Posted by : admin / Posted on : May 17, 2021

        1) When measuring mixed-phase fluids, choose a place that will not cause phase separation. When measuring two-component liquids, avoid installing them in the downstream where the mixing is not uniform. When measuring chemical reaction pipelines, install them in the fully completed section of the reaction. 
  2) Avoid the negative pressure in the measuring tube as much as possible. 
  3) Choose a place with low vibration, especially for integrated instruments.
  4). Avoid large motors, large transformers, etc. nearby to avoid electromagnetic field interference. 
  5) A place where the sensor is easily grounded. 
  6) Avoid high-concentration corrosive gases in the surrounding environment as much as possible. 
  7) The ambient temperature is within the range of -25-60℃, and the relative humidity of the environment is within the range of 10%-9O%. Avoid direct sunlight as much as possible. 
     8) The liquid should have the conductivity required for measurement, and the conductivity distribution should be generally uniform. Therefore, the installation of the flowmeters sensor should avoid places that are prone to uneven conductivity, such as adding liquid medicine near its upstream, and the liquid filling point is best located downstream of the sensor.